Development of a cone-shape phantom for multi-voxel MR spectroscopy

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

0
10
0

Year Published

2008
2008
2018
2018

Publication Types

Select...
5
1

Relationship

1
5

Authors

Journals

citations
Cited by 9 publications
(10 citation statements)
references
References 33 publications
0
10
0
Order By: Relevance
“…This factor can be measured as the signal (initially) versus noise (after full relaxation) of the free induction decay (FID) spectrum in the time domain, or as the height of the largest metabolite peak divided by the noise in the frequency domain. The SNR of the metabolite peak depends on the metabolite concentrations (11,12,16,17); however, the concentration of water ($55.6 mM) in various MR phantoms is almost identical. (14) Therefore, the water resonance can be used as a reference for SNR calculations in different phantoms.…”
Section: Principal Quality Factors For Mrs Qa/qc Snrmentioning
confidence: 99%
See 3 more Smart Citations
“…This factor can be measured as the signal (initially) versus noise (after full relaxation) of the free induction decay (FID) spectrum in the time domain, or as the height of the largest metabolite peak divided by the noise in the frequency domain. The SNR of the metabolite peak depends on the metabolite concentrations (11,12,16,17); however, the concentration of water ($55.6 mM) in various MR phantoms is almost identical. (14) Therefore, the water resonance can be used as a reference for SNR calculations in different phantoms.…”
Section: Principal Quality Factors For Mrs Qa/qc Snrmentioning
confidence: 99%
“…We attempted to measure the localization and quantification accuracy by evaluating proportional relationships between the signal intensity of a metabolite and the quantity of the metabolite. A cone-shaped phantom was introduced in a previous study (16) as a possible tool for the evaluation of MR signals using the same concept. However, the use of this phantom in a previous study was confounded by air bubbles that caused susceptibility effects (55) in the cone-shaped vials due to a single opening to the inlet solution.…”
Section: Long-term Testmentioning
confidence: 99%
See 2 more Smart Citations
“…6,7 Detail phantoms utilizing acrylic or glass containers are vulnerable to susceptibility artifacts arising from the interface of the compartment wall material and the solution used to fill the phantom. 8 Those artifacts are more pronounced in high-detail and irregularly shaped phantoms.…”
Section: Introductionmentioning
confidence: 99%