Development of an Accurate Resonant Frequency Controlled Wire Ultrasound Surgical Instrument

Abstract: Our developed wire ultrasound surgical instrument comprises a bolt-clamped Langevin ultrasonic transducer (BLUT) fabricated by PMN-PZT single crystal material due to high mechanical quality factor and electromechanical coupling coefficient, a waveguide in the handheld instrument, and a generator instrument. To ensure high performance of wire ultrasound surgical instruments, the BLUT should vibrate at an accurate frequency because the BLUT’s frequency influences hemostasis and the effects of incisions on blood … Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
13
0

Year Published

2020
2020
2022
2022

Publication Types

Select...
10

Relationship

7
3

Authors

Journals

citations
Cited by 15 publications
(13 citation statements)
references
References 42 publications
0
13
0
Order By: Relevance
“…As a result, the outputs of the amplifier and BWE circuit-equipped amplifier were measured and compared. The voltage gain is a performance indicator for measuring the performance of an amplifier, and its high output helps to provide a clear ultrasound image [ 44 , 45 , 46 ]. In addition, the bandwidth of the gain over frequency of the amplifier only and the amplifier equipped with the BWE circuit can be compared.…”
Section: Methodsmentioning
confidence: 99%
“…As a result, the outputs of the amplifier and BWE circuit-equipped amplifier were measured and compared. The voltage gain is a performance indicator for measuring the performance of an amplifier, and its high output helps to provide a clear ultrasound image [ 44 , 45 , 46 ]. In addition, the bandwidth of the gain over frequency of the amplifier only and the amplifier equipped with the BWE circuit can be compared.…”
Section: Methodsmentioning
confidence: 99%
“…Additionally, the pulse signal from the dynamic pulser has certain frequency characteristics, and a feedback route can be formed through the capacitor to block the DC signal used in the amplifier; thus, the distributed signal is minimized and the feedback route is blocked by using CC1 and high values of RC1. RC2 and RC3 are appropriately selected to have the reference voltage V REF to ensure stable operation and prevent oscillation [ 37 ]. V GG is used to provide stable DC voltage used in the comparator, dynamic pulser, and gate of the power amplifier.…”
Section: Dynamic Bias Technique For Ultrasonic Transmittersmentioning
confidence: 99%
“…The BVD circuit model is shown in Figure 1a. The value of the circuit elements can be derived from the simulated or measured impedance response of the design by solving the following expressions for the resonance f r and anti-resonance f a frequencies [16,17]:…”
Section: Performance Metricsmentioning
confidence: 99%