2018
DOI: 10.1063/1.4993928
|View full text |Cite
|
Sign up to set email alerts
|

Development of ball surface acoustic wave trace moisture analyzer using burst waveform undersampling circuit

Abstract: The measurement and control of trace moisture, where the water concentration is lower than 1 ppmv [-76.2 °C for the frost point (°CFP)], are essential for improving the yield rate of semiconductor devices and for ensuring their reliability. A ball surface acoustic wave (SAW) sensor with a sol-gel silica coating exhibited useful characteristics for a trace moisture analyzer (TMA) when the temperature drift of the delay time output was precisely compensated using two-frequency measurement (TFM), where the temper… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
10
0

Year Published

2019
2019
2024
2024

Publication Types

Select...
7

Relationship

2
5

Authors

Journals

citations
Cited by 9 publications
(10 citation statements)
references
References 27 publications
0
10
0
Order By: Relevance
“…41) Four waveforms of 16 bit with a length of 300 points were simultaneously recorded at a sampling rate of 100 MHz using BUS circuit, representing the signals of 80 and 240 MHz at the third turn and the seventh turn. 45) They were recorded every 3 s at a minimum interval under the current BUS protocol, and then the maximum averaging factor was 1000.…”
Section: Measurement Systemmentioning
confidence: 99%
See 3 more Smart Citations
“…41) Four waveforms of 16 bit with a length of 300 points were simultaneously recorded at a sampling rate of 100 MHz using BUS circuit, representing the signals of 80 and 240 MHz at the third turn and the seventh turn. 45) They were recorded every 3 s at a minimum interval under the current BUS protocol, and then the maximum averaging factor was 1000.…”
Section: Measurement Systemmentioning
confidence: 99%
“…As R y decreased, t s decreased, where t s was 60, 40, and 15 s for the BA, EP, and ECB tubes, respectively. We succeeded in measuring the transient behavior of the ECB tube for the first time by reducing the measurement interval from 14.2 45,47) to 3 s. There was a significant difference in t s between EP and ECB treatments despite the length being only 100 mm. This result implies that the selection of the inner surface treatment is important for constructing highly pure gas distribution systems.…”
Section: Moisture Adsorptionmentioning
confidence: 99%
See 2 more Smart Citations
“…We have developed a ball SAW TMA with the twofrequency measurement (TFM) of delay time to compensate for the variation in temperature 14,15) as well as to measure the temperature. 16) In the FP range of −60 to −20 °C, the attenuation of a SAW rather than its velocity is used to measure the moisture content.…”
mentioning
confidence: 99%