A robust, rapid and sensitive UPLC-MS/MS method has been developed, optimized and validated for the determination of amlodipine (AML) and atorvastatin (ATO) in human plasma using eplerenone as an internal standard (IS). Multiple-reaction monitoring in positive electrospray ionization mode was utilized in Xevo TQD LC-MS/MS. Double extraction was used in sample preparation using diethyl ether and ethyl acetate. The prepared samples were analyzed using an Acquity UPLC BEH C (50 × 2.1 mm, 1.7 μm) column. Ammonium formate and acetonitrile, pumped isocraticaly at a flow rate of 0.25 mL/min, were used as a mobile phase. Method validation was done as per the US Food and Drug Administration guidelines. Linearity was achieved in the range of 0.1-10 ng/mL for AML and 0.05-50 ng/mL for ATO. Intra-day and inter-day accuracy and precision were calculated and found to be within the acceptable range. A short run time, of <1.5 min, permits analysis of a large number of plasma samples per batch. The developed and validated method was applied to estimate AML and ATO in a bioequivalence study in healthy human volunteers.