Background: HemoTypeSC is a rapid, point-of-care testing (POCT) device for sickle cell disease (SCD) that traditionally uses the capillary blood from heel stick collected at the point of testing, a procedure that makes mass screening cumbersome and less cost-effective. Using dried blood spots (DBS) on HemoTypeSC could mitigate this challenge. Therefore, this study aimed to determine the feasibility of eluting blood from DBS to read on HemoTypeSC.Methods: DBS and fresh samples from heel sticks were collected from 511 newborns at the immunization clinics of six Primary Health Centers in Abuja, Nigeria. The two samples from each newborn were analyzed using HemoType SC and then compared with the result of the isoelectric focusing (IEF) test.Results: Of the 511 newborns, 241 were males and 270 were females. Standard HemoTypeSC (using fresh samples collected from heel sticks) and HemoTypeSC using DBS identified 404 (79.0%) HbAA, 100 (19.6%) HbAS, 6 (1.2%) HbSS, and 1 (0.2%) HbAC phenotypes. The IEF tests identified 370 (72.4%) HbAA, 133 (26.0%) HbAS, 5 (1.0%) HbSS, and 3 (0.6%) HbAC phenotypes. The sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) and overall accuracy of HemoTypeSC using DBS, compared to standard HemoTypeSC POCT was 100%. IEF method showed for AA, AS, AC phenotypes; sensitivity; 84.7%, 67%,100% respectively, specificity; 67.6%, 86%, 99% respectively, PPV; 91.2%, 53%, 50% respectively, NPV; 52.7%, 91%, 100% respectively. For SS phenotype, IEF showed 100% specificity, sensitivity, PPV and NPV.Conclusion: HemoTypeSC test using dried blood spot is as accurate as the standard point-of-care HemoTypeSC test. The use of DBS on HemoTypeSC could ensure better efficiency and cost-effectiveness in mass newborn screening for SCD.