2009
DOI: 10.1016/j.wear.2009.01.028
|View full text |Cite
|
Sign up to set email alerts
|

Differences in wear between load and displacement control tested total knee replacements

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

2
21
0

Year Published

2009
2009
2022
2022

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 21 publications
(23 citation statements)
references
References 5 publications
2
21
0
Order By: Relevance
“…The mean wear rate of 18 mm 3 per year for the nondelaminated tibial liners falls in the range of 9 and 23 mm 3 per million cycles obtained in physiologic knee simulators on the same or similar TKA designs [27,32,35] (Table 3), where 1 million cycles is considered to be approximately equivalent to 1 year in vivo. Other than the study by Knowlton and Wimmer [19], we are aware of only one other study reporting measured volumetric liner wear in retrieved TKRs [8], and it is for contemporary prosthesis designs (LCS 1 , P.F.C.…”
Section: Discussionmentioning
confidence: 86%
“…The mean wear rate of 18 mm 3 per year for the nondelaminated tibial liners falls in the range of 9 and 23 mm 3 per million cycles obtained in physiologic knee simulators on the same or similar TKA designs [27,32,35] (Table 3), where 1 million cycles is considered to be approximately equivalent to 1 year in vivo. Other than the study by Knowlton and Wimmer [19], we are aware of only one other study reporting measured volumetric liner wear in retrieved TKRs [8], and it is for contemporary prosthesis designs (LCS 1 , P.F.C.…”
Section: Discussionmentioning
confidence: 86%
“…The current study is in good agreement with wear simulator studies of similar TKR designs. Schwenke et al tested two groups of cruciate‐retaining MGII components with ISO displacement‐ and load‐controlled level walking and found steady‐state wear rates of 9.2 ± 0.9 mg/Mc and 20.9 ± 4.2 mg per million cycles, respectively . CMM measurements of a similar fixed bearing cruciate retaining design run for 5 million cycles of ISO load‐controlled level walking found a wear rate of 8.3 ± 0.9 mm 3 /million cycles and a creep volume loss of 36 ± 4 mm 3 .…”
Section: Discussionmentioning
confidence: 99%
“…The averaged motion data per station were then averaged to obtain overall motion profiles. Further details of the wear test are reported in Schwenke et al 2009 [16]. …”
Section: 0 Materials and Methodsmentioning
confidence: 99%