a b s t r a c tOzone (O 3 ) flux (F o ) was measured over a wheat field in Yucheng, China, using the eddy covariance technique. A dry chemiluminescence fast-response O 3 analyzer and a UV-absorption based slow-response O 3 analyzer were used for measuring fast O 3 concentration fluctuations and absolute concentrations, respectively. The main objective of this study was to determine a group of suitable methods for calculating F o based on the performance of the two O 3 analyzers in the field. We evaluated the effects of three calibration methods on F o . These calibration methods include the ratio method (RM, it assumes that the fast-response analyzer's signal-output is directly proportional to absolute ambient O 3 concentration within a 30-min interval), the ratio offset method (ROM, it is based on the ratio method with analyzer's offset modification), and the ratio variation method (RVM, it is based on the ratio method with the fastresponse analyzer's gain factor variation within a 30-min interval). Three frequency response corrections (analytic, revised analytic, and in-situ) for the estimation of F o were also evaluated. Our results show that: (1) Compared to the O 3 flux with the commonly-used method (RM), on average, the fluxes with ROM and RVM were decreased by about 9% and 7%, respectively. (2) Based on the spectral analysis, the fastresponse O 3 analyzer's noise could cause random errors, but did not result in a systematic error in F o . Tubing attenuation in O 3 concentration fluctuations can cause up to 20% loss in F o . (3) Frequency response corrections for F o with the original analytic method, the in-situ method, and the revised analytic method were 34.6%, 23.8% and 25.4% on average, respectively.