Disposable Airway Pressure Manometers for Endotracheal Tube Cuff Inflation

Abstract: This study aimed to assess the performance, accuracy, precision and repeatability of two single-use airway pressure manometers as a cost-effective alternative for inflation of endotracheal tubes with high-volume, low-pressure cuffs. The manometers were tested in a bench top model against a U-tube manometer. Eighteen units of each device were tested. Three consecutive measurements were performed at pressures of 20, 25 and 30 cmH2O each. The mean ± SD of the recorded pressures and maximum deviation from the targ… Show more