Arterial compliance is one of the major indices characterizing arterial stiffness. Electrical analogy artery model, which incorporates elements, characterizing all four main effects of the blood flow, and enables evaluation of pressure-dependent compliance using non-invasive techniques, was proposed in this study. Assumption that, blood inertance together with non-linear compliance enables more accurate representation and provides quantitatively and qualitatively improved diagnostic information, was validated in initial experimental in vivo study of human carotid artery. Mean value of inadequancy for C P -9,69 % was obtained comparing proposed four and two element models. Ill. 5, bibl. 6, tabl. 3 (in English; abstracts in English and Lithuanian). I. Kupčiūnas, A. Kopustinskas. Elektrinės analogijos arterijos modelis netiesiniam kompliansui įvertinti // Elektronika ir elektrotechnika. -Kaunas: Technologija, 2012. -Nr. 1(117). -P. 95-98.Arterijos kompliansas yra vienas iš svarbiausių parametrų, apibūdinančių arterijų standumą. Sudarytas elektrinės analogijos arterijos modelis atspindi visus keturis pagrindinius kraujotakos procesus ir leidžia neinvaziniais metodais įvertinti nuo slėgio priklausantį kompliansą. Prielaida, kad netiesinio komplianso ir kraujo inertiškumo įvertinimas įgalina tiksliau apibūdinti kraujotaką ir gauti papildomą kokybinę ir kiekybinę diagnostinę informaciją, patvirtinta atlikus in vivo žmogaus karotidinės arterijos pradinį eksperimentinį tyrimą. Nustatyta vidutinė 9,69 % C P verčių neatitiktis pasiūlytiems keturių ir dviejų elementų modeliams. Il. 5, bibl. 6, lent. 3 (anglų kalba; santraukos anglų ir lietuvių k.).