Dosimetric Impact of the Positional Imaging Frequency for Hypofractionated Prostate Radiotherapy – A Voxel-by-Voxel Analysis

Abstract: Background To investigate deviations between planned and applied treatment doses for hypofractionated prostate radiotherapy and to quantify dosimetric accuracy in dependence of the image guidance frequency. Methods Daily diagnostic in-room CTs were carried out in 10 patients in treatment position as image guidance for hypofractionated prostate radiotherapy. Fraction doses were mapped to the planning CTs and recalculated, and applied doses were accumulated voxel-wise usi… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
4
0

Year Published

2021
2021
2023
2023

Publication Types

Select...
4
1

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(4 citation statements)
references
References 43 publications
0
4
0
Order By: Relevance
“…As strict adherence to organ preparation protocols is still mandatory, selecting the closest temporal CBCT in case of missing CBCTs is expected to have no excessive impact on the dose accumulation accuracy. Nonetheless, daily instead of non-daily CBCT may further improve the dose accumulation workflow and will be explored in sequential studies [ 20 , 21 ]. Our data is also limited by the absence of intra-fractional motion analysis and correction, which was not available for the analyzed patient cohort.…”
Section: Discussionmentioning
confidence: 99%
“…As strict adherence to organ preparation protocols is still mandatory, selecting the closest temporal CBCT in case of missing CBCTs is expected to have no excessive impact on the dose accumulation accuracy. Nonetheless, daily instead of non-daily CBCT may further improve the dose accumulation workflow and will be explored in sequential studies [ 20 , 21 ]. Our data is also limited by the absence of intra-fractional motion analysis and correction, which was not available for the analyzed patient cohort.…”
Section: Discussionmentioning
confidence: 99%
“…Images were transferred from the Eclipse 15.6 treatment planning system to the Velocity software V4.0 (Varian Medical Systems, Palo Alto, CA) for image registration and dose reconstruction. The accuracy of deformable image registration and dose reconstruction using computer simulations and patient data has been demonstrated in previous papers, and has been tested and validated separately [18,19] . The clinical shift was used for rigid alignment between partial KVCBCT images and the corresponding planned CT.…”
Section: Methodsmentioning
confidence: 99%
“…Images were transferred from the Eclipse 15.6 treatment planning system to the Velocity software V4.0 (Varian Medical Systems, Palo Alto, CA) for image registration and dose reconstruction. The accuracy of deformable image registration and dose reconstruction using computer simulations and patient data has been demonstrated in previous papers, and has been tested and validated separately [ 20 , 21 ]. The clinical shift was used for rigid alignment between partial KVCBCT images and the corresponding planned CT.…”
Section: Methodsmentioning
confidence: 99%