“…The Attune component showed a 2 mm radial press fit with maximum incomplete seating of 1.8 mm for the cement mantle instrumentation. The radial press fit effect as reported by Jaeger et al 38 could be enhanced in sclerotic bone. Evaluation of the postoperative radiographs showed no significant differences in the cement penetration area for the AP or the lateral radiographs between both investigated groups.…”
Section: Discussionsupporting
confidence: 60%
See 1 more Smart Citation
“…The Attune component showed a 2 mm radial press fit with maximum incomplete seating of 1.8 mm for the cement mantle instrumentation. The radial press fit effect as reported by Jaeger et al 38 could be enhanced in sclerotic bone. Evaluation of the postoperative radiographs showed no significant differences in the cement penetration area for the AP or the lateral radiographs between both investigated groups.…”
Section: Discussionsupporting
confidence: 60%
“…The ligament tension force during cement hardening and the impaction force must exceed the radial press fit. Jaeger et al 38 found an increased radial press fit around the tibial stem and keel, with growing radii of the rounded tibial edges and with decreasing cement mantle space around the implants. The Attune component showed a 2 mm radial press fit with maximum incomplete seating of 1.8 mm for the cement mantle instrumentation.…”
Section: Discussionmentioning
confidence: 98%
“…A recent report showed that excessive press fit may lead to incomplete seating or tilting of the tibial component, especially in hard and uneven sclerotic bone. 52 Another factor attributed to tibial loosening is stress shielding. The ATTUNE system uses a thick CoCr tibial baseplate, and there has been a report of an increased rate of medial tibial bone resorption with the ATTUNE TKA, presenting in various locations and severities around the baseplate.…”
Section: Discussionmentioning
confidence: 99%
“…Regarding the latter, per manufacturer literature, the surface roughness of the tibial component is decreased compared to the femoral component (60 grit blasts vs. 220 grit blasts). Compared to the femoral component, the decreasing grit blast combined with the increased rotational forces and reduced surface area of the tibial baseplate may lead to the failures seen at the tibial implant-bone cement interface [ 4 ].…”
Section: Discussionmentioning
confidence: 99%