Residual respiratory motion degrades image quality in conventional cardiac cine MRI (CCMR). We evaluated whether a free-breathing (FB) radial imaging CCMR sequence with compressed sensing reconstruction (eXtra-Dimension (e.g. cardiac and respiratory phases) Golden-angle RAdial Sparse Parallel, or XD-GRASP) could provide better image quality than a conventional Cartesian breath-held (BH) sequence, in an unselected population of patients undergoing clinical CCMR.
Material and Methods
101 patients who underwent BH and FB imaging in a mid-ventricular short-axis plane at a matching location were included. Visual and quantitative image analysis was performed by two blinded experienced readers, using a 5-point qualitative scale to score overall image quality and visual signal-to-noise ratio (SNR) grade, with measures of noise and sharpness. End-diastole (ED) and end-systole (ES) left-ventricular areas were also measured and compared for both BH and FB images.
Results
Image quality was generally better with the BH cines (overall quality grade BH vs FB: 4 vs 2.9, p<0.001; noise 0.06 vs 0.08 p< 0.001; SNR grade: 4.1 vs 3, p<0.001), except for sharpness (p=0.48). There were no significant differences between BH and FB images regarding ED or ES areas (p=0.35 and 0.12). 18 of the 101 patients had impaired BH image quality (grades 1 or 2). In this subgroup, image quality of the FB images was better (p=0.0032), as was the SNR grade (p=0.003), but there were no significant differences regarding noise and sharpness (p=0.45, p=0.47).
Conclusion
Although FB XD-GRASP CCMR was visually inferior to conventional BH cardiac cine in general, it provided improved image quality in the subgroup of patients presenting respiratory motion-induced artifacts on breath-held images.