Effect of infusion line connection and infusion line characteristics on start‐up fluid delivery from syringe infusion pumps used for microinfusion
Markus Weiss,
Pedro David Wendel‐Garcia,
Vincenzo Cannizzaro
et al.
Abstract:BackgroundConnecting an infusion line to a closed stopcock results in pressurization of fluid within the syringe infusion pump assembly leading to flow irregularities when opening the stopcock and activating the pump.AimsIt was the purpose of this study to assess the extent of pressurization under different conditions and its impact on start‐up fluid delivery.MethodsIntraluminal pressures and start‐up fluid delivery at 1 mL/h flow rate were assessed with connection of the infusion line 1 min (delayed connectio… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.