The results of this study provide important insight into the relation between NRT thresholds and loudness judgments for different stimulation rates and T- and C-levels at various MAP rates. The loudness judgment dynamic range and MAP dynamic range (T- and C-levels) varied notably for different stimulation rates. As a result, the relation of NRT thresholds to these measures also varied with stimulation rate. Overall, the mean vNRT thresholds fell higher in the loudness judgment dynamic range than the tNRT thresholds. Mean NRT thresholds fell between the judgments of medium soft and maximum acceptable loudness for all stimulation rates. Mean vNRT thresholds fell above C-levels, whereas almost half of tNRT thresholds fell just below C-levels. However, the relation between NRT thresholds and C-levels varied substantially for different MAP stimulation levels. In addition, there is substantial individual variability in the relation between NRT thresholds and MAP levels that is not reflected in the group data. The prediction of the contour of T- and C-levels from the contour of NRT thresholds across electrodes would not be appropriate for half of the subjects. Therefore, great care should be taken when applying a fitting rule that incorporates NRT thresholds without considering these individual differences. For adults who can provide appropriate loudness judgments and threshold responses it appears to be most efficient to primarily use behavioral measures to create MAPs.