2016
DOI: 10.2298/vsp141208041m
|View full text |Cite
|
Sign up to set email alerts
|

Effect of surgical drill guide and irrigans temperature on thermal bone changes during drilling implant sites - thermographic analysis on bovine ribs

Abstract: Guided implant site preparation generates higher temperature of the local bone than conventional drilling, not exceeding the threshold for thermal bone necrosis. Although saline at room temperature provides sufficient heat control during drilling, cooled saline is more effective regardless the use of surgical drill guide.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4

Citation Types

3
24
1
2

Year Published

2019
2019
2023
2023

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 26 publications
(30 citation statements)
references
References 17 publications
3
24
1
2
Order By: Relevance
“…During guided sequential drilling with a 4.2 mm final drill, the time over critical temperature (time with a temperature increase over 15°C for a 25°C polyurethane foam block) was more than 1 min and exceeded the threshold for causing morphological bone damage (Eriksson & Albrektsson, ). Our results do not correspond to the findings of other in vitro studies which documented that heat generation with GOP did not reach temperature levels critical to bone (Dos Santos et al, ; Liu et al, ; Markovic et al, ; Migliorati et al, ; Misir et al, ).…”
Section: Discussioncontrasting
confidence: 99%
See 4 more Smart Citations
“…During guided sequential drilling with a 4.2 mm final drill, the time over critical temperature (time with a temperature increase over 15°C for a 25°C polyurethane foam block) was more than 1 min and exceeded the threshold for causing morphological bone damage (Eriksson & Albrektsson, ). Our results do not correspond to the findings of other in vitro studies which documented that heat generation with GOP did not reach temperature levels critical to bone (Dos Santos et al, ; Liu et al, ; Markovic et al, ; Migliorati et al, ; Misir et al, ).…”
Section: Discussioncontrasting
confidence: 99%
“…Higher temperature changes were observed in GOP compared to CA, and in sequential compared to single drilling. This is in line with several other studies that suggest the greater heat generation with guided procedures is caused by the surgical guide impairing the irrigation fluid from entering the drilling site (Dos Santos et al, ; Markovic et al, ; Migliorati et al, ; Misir et al, ). Drilling with a cooling channel in the guide has been shown to reduce the temperature increase by 1.9‐fold (Liu et al, ).…”
Section: Discussionsupporting
confidence: 89%
See 3 more Smart Citations