Effects of Environmental and Mechanical Variations on Particle Size Distribution of Albuterol HFA Inhalers

Abstract: BACKGROUND:The purpose of this study was to evaluate how factors (ambient temperature, shaking the inhaler before use, suspension of the inhaler in water, and the variation over the lifetime of the inhaler) affect the particle-size distributions from albuterol HFA inhalers. METHODS: We used a laser particle-size analyzer to measure the percentage of particles in the 1-to 5-lm range (fine-particle fraction) serially 2,500 times per second to obtain a window of useful measurements with each inhaler actuation. We… Show more