12 Open Journal of Optimization tion of the peaks and baseline. The progressive optimization of different parameters such as the stationary phase (C18 YMC 150 mm * 4.6 mm -3 µm), the proportion of the mobile phase: acetonitrile/water (80/20), the flow rate impelled by the pump (0.8 mL/min) and the modification of the analyte preparation mode (same amount of PHG and TPHG in a 50 mL volumetric flask) resulted in a final chromatogram that highlighted two chromatographic peaks at the respective RT of 2.391 min (PHG) and 3.735 min (TPHG) at a run time of 6 min. The chromatographic conditions that led to the final chromatogram can be used routinely by the pharmaceutical industry for the determination of several PHG and TPHG drug matrices after prior validation of the determination method.