Eliminating Mo isobaric interference using O2 as reaction gas for Tc measurement by triple quadrupole ICP-MS

Abstract: Monitoring of ultra-trace level 99Tc in environmental samples is important for radioactivity assessment and tracing oceanographic processes. 97Tc was an ideal yield tracer for the analysis of 99Tc. However, when...

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

2022
2022
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 6 publications
(1 citation statement)
references
References 23 publications
0
1
0
Order By: Relevance
“…In the ICP-MS/MS determination of radioisotopes using a commercial high-efficiency desolvating nebuliser, the 97 Mo isobaric interference on 97 Tc was eliminated 92 by adding O 2 as a reaction gas to form MoO + and MoO 2 + . Using 97 Tc as a yield tracer for 99 Tc, the absolute LOD was 0.9 fg (0.6 mBq).…”
Section: Water Analysismentioning
confidence: 99%
“…In the ICP-MS/MS determination of radioisotopes using a commercial high-efficiency desolvating nebuliser, the 97 Mo isobaric interference on 97 Tc was eliminated 92 by adding O 2 as a reaction gas to form MoO + and MoO 2 + . Using 97 Tc as a yield tracer for 99 Tc, the absolute LOD was 0.9 fg (0.6 mBq).…”