Objective: To investigate the potential of 2D ion chamber arrays to serve as a standalone tool for the verification of source strength, positioning and dwell time, within the framework of 192Ir high-dose rate brachytherapy device quality assurance. Approach: A commercially available ion chamber array was used. Fitting of a 2D Lorentzian peak function to experimental data from a multiple source dwell position irradiation on a frame-by-frame basis, facilitated tracking of the source center orthogonal projection on the array plane. For source air kerma strength verification, Monte Carlo simulation was employed to obtain a chamber array- and source-specific correction factor of calibration with a 6 MV photon beam. This factor converted the signal measured by each ion chamber element to air kerma in free space. A source positioning correction was also applied to lift potential geometry mismatch between experiment and Monte Carlo simulation. Main Results: Spatial and temporal accuracy of source movement was verified within 0.5 mm and 0.02 s, respectively, in compliance with the test endpoints recommended by international professional societies. The source air kerma strength was verified experimentally within method uncertainties estimated as 1.45% (k=1). The source positioning correction method employed did not introduce bias to experimental results of irradiations where source positioning was accurate. Development of a custom jig attachable to the chamber array for accurate and reproducible experimental set up would improve testing accuracy and obviate the need for source positioning correction in air kerma strength verification. Significance: Delivery of a single irradiation plan, optimized based on results of this work, to a 2D ion chamber array can be used for concurrent testing of source position, dwell time and air kerma strength, and the procedure can be expedited through automation. Chamber arrays merit further study in treatment planning quality assurance and real time, in vivo dose verification.