Abstract:We report on a simple method to perform an absolute calibration of the magnetic beam analyzer of the reaccelerator ReA3 at the National Superconducting Cyclotron Laboratory. The method is based on the time of flight between two beam stoppers 7.65 m apart. Based on two independent time-of-flight measurements at three different beam energies, the beam analyzer magnet is calibrated with an accuracy of 0.12 %, corresponding to a beam energy accuracy of 0.24 %.
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.