Energy response of X-rays under high flux conditions using a thin APD for the energy range of 6–33 keV

Abstract: This paper reports on the demonstration of a high-rate energy measurement technique using a thin depletion layer silicon avalanche photodiode (Si-APD). A dedicated amplitude-to-time converter is developed to realize simultaneous energy and timing measurement in a high rate condition. The energy response of the system is systematically studied by using monochromatic X-ray beam with an incident energy ranging from 6 to 33 keV. The obtained energy spectra contain clear peaks and tail distributions. The peak fract… Show more

Help me understand this report
View preprint versions

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
4
0

Year Published

2019
2019
2021
2021

Publication Types

Select...
3
2

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(4 citation statements)
references
References 31 publications
0
4
0
Order By: Relevance
“…The conceptual diagram of the signal processing is shown in Fig 6 . Each Si-APD output is amplified by a fast amplifier (Mini-circuits, RAM-8A+) located directly behind the Si-APDs. The amplified output is processed in several ways to obtain event characteristics 19,20 . First of all, it is converted to a logic signal by a lab-built constant fraction discriminator (CFD); this yields an event occurrence time.…”
Section: Appendices a Detector System A1 Apd Detector And Readout Systemmentioning
confidence: 99%
See 1 more Smart Citation
“…The conceptual diagram of the signal processing is shown in Fig 6 . Each Si-APD output is amplified by a fast amplifier (Mini-circuits, RAM-8A+) located directly behind the Si-APDs. The amplified output is processed in several ways to obtain event characteristics 19,20 . First of all, it is converted to a logic signal by a lab-built constant fraction discriminator (CFD); this yields an event occurrence time.…”
Section: Appendices a Detector System A1 Apd Detector And Readout Systemmentioning
confidence: 99%
“…The values in the parentheses are those normalized to the sum, which is shown in the bottom row. The energy dependence of ε apd , together with the APD response function, is studied by a separate experiment at KEK-PF 19 ; it is expressed by ε apd (E) 1 − exp[−σ Si (E)ρ Si L Si ], where σ Si (E) is the photoelectric absorption cross section 26 of Si at the energy E, ρ Si the density and L Si an effective thickness of the Si sensor. Some uncertainties exit in determining L Si , but their effect is negligible for the ratio calculation.…”
Section: B22mentioning
confidence: 99%
“…The pellet was covered with two 20 Â 20 mm MgF 2 substrates, one on either side. The scattered X-ray photons were detected by a dedicated energy-sensitive X-ray detector (Masuda et al, 2017(Masuda et al, , 2019a.…”
Section: Nrs Measurementmentioning
confidence: 99%
“…The pellet was covered with two 20 × 20 mm 2 MgF 2 substrates. The scattered X-ray photons were detected by a dedicated energy-sensitive X-ray detector [25,26].…”
Section: Nrs Measurementmentioning
confidence: 99%