Enteral stent construction: Current principles

Abstract: A B S T R A C TThe insertion of self-expanding stents into malignant strictures of the small and large bowel has become a routine procedure around the world. However, stent development has happened very much on a "trial & error" approach, based mostly on bright ideas of enthusiastic individuals or marketing decisions by the manufacturer. A large variety of stents are commercially available, covered to a variable degree by a membrane to reduce tissue ingrowth. However, in vitro characteristics and in vivo behav… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1

Citation Types

0
4
0

Year Published

2022
2022
2023
2023

Publication Types

Select...
3

Relationship

1
2

Authors

Journals

citations
Cited by 3 publications
(4 citation statements)
references
References 33 publications
0
4
0
Order By: Relevance
“…The stent is available in lengths varying from 8 to 15 cm and a body diameter of 20 mm with flanges of 25 mm and is placed with a pistol-grip delivery system which allows controlled release and recapturing, if needed. The braided construction and the silicone encasement of the stent mesh allows for a higher expansion force than the Ultraflex ® , making it ideal for placement in strictures with a large tumor bulk or surrounding fibrosis in the mid-esophagus [26, 45] (Fig. 4).…”
Section: Stent Selectionmentioning
confidence: 99%
See 2 more Smart Citations
“…The stent is available in lengths varying from 8 to 15 cm and a body diameter of 20 mm with flanges of 25 mm and is placed with a pistol-grip delivery system which allows controlled release and recapturing, if needed. The braided construction and the silicone encasement of the stent mesh allows for a higher expansion force than the Ultraflex ® , making it ideal for placement in strictures with a large tumor bulk or surrounding fibrosis in the mid-esophagus [26, 45] (Fig. 4).…”
Section: Stent Selectionmentioning
confidence: 99%
“…In the braided stents, the nitinol wires cross over each other but do not interlock. They have a high radial force and, although being more flexible, that configuration retains a relativity high axial force specially when coated, due to the restriction of the movement of the wires by the silicone dipping, heightening the risk of kinking on flexion [26]. In addition, braided stents shorten significantly upon expansion, thus being less predictable when deployed.…”
Section: Stent Materials and Constructionmentioning
confidence: 99%
See 1 more Smart Citation
“…7a). Braided stents have a high radial expansive force whilst also retaining flexibility [71]. The wires are free to move across each other at their points of contact, allowing the stent to conform during muscle contractions in the oesophagus [33,72].…”
Section: Braided Stentsmentioning