Objective:The aims of this study were to compare the central corneal thickness (CCT) measured by an ultrasonic pachymeter (SP-100, Tomey, Japão) versus the Galilei G6™ (Ziemer Ophthalmics System AG, Port, Switzerland) and to compare the simulated keratometry (SimK) measured by an autokeratometer (KM 500 Nidek) versus the Galilei G6™.Animal studied: Eighteen mixed-breed cats, 10 males, and eight females, aged between 18 and 48 months, were evaluated (n = 36 eyes).
Procedure(S):The cats were manually restrained using a blanket. The operator held the autokeratometer close to the eye and measured the SimK. Next, one anesthetic eye drop was applied; the operator lightly touched the corneal using the ultrasound pachymeter and measured the CCT. At another moment, the cats were sedated using meperidine and xylazine and then evaluated using the Galilei.Student's t-test was employed to perform the statistical analyzes.Results: Considering the CCT, the ultrasound pachymeter provided mean ± SD of 623.03 ± 48.17 μm and the Galilei 617.34 ± 53.38 μm. The autokeratometer produced a SimK of 38.20 ± 0.84D (8.84 ± 0.20 mm) and the Galilei 38.37 ± 0.83D