Abstract:Narrow spectrum series from N-40 to N-120 were established according to ISO 4037:2019. This study was made through experimental measurements and the software modeling toolkit SpekPy v2.0. Experiments were performed to determine Inherent filtration (0.332 & 0.492 mm Al at 100 cm & 250 cm, respectively). The maximum deviation compared to ISO 4037:2019 of the measured HVLs values are not more then 0.10 mm Al and for modeling not exceed 0.09 mm Al; h range 0.91 to 0.97 for experimently and 0.89 to 0.96 for… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.