Estimates of Patient Radiation Doses in Digital Radiography Using DICOM Information at a Large Teaching Hospital in Oman

Abstract: In this study, we sought to estimate the patient radiation doses in the digital radiography X-ray examinations conducted in a large hospital. The patient exposure factors and kerma-area product (P KA ) were retrospectively recorded via the Digital Imaging and Communications in Medicine (DICOM) header for 547 patients. The entrance surface air kerma (ESAK) was estimated from the measurements of the X-ray tube output and recorded exposure factors, as well as from the console that displayed P KA as an alternative… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1
1
1

Citation Types

2
13
0
2

Year Published

2020
2020
2024
2024

Publication Types

Select...
7
1
1

Relationship

1
8

Authors

Journals

citations
Cited by 23 publications
(17 citation statements)
references
References 19 publications
2
13
0
2
Order By: Relevance
“…Table 6 shows a summary of this study’s LDRL values that they were lower than those of international organisations (EC 2018 ; IAEA 2004 ) and other previously published studies (Ahmed et al 2020 ; Baş Mor et al 2018 ; Hart et al 2012 ; Hoseini Motlagh et al 2020 ; Metaxas et al 2019 ; Mohsenzadeh et al 2018 ; Nyathi et al 2009 ; Suliman 2020 ). The reason for this is that the radiographers at the research site used the high kVp and low mAs technique to produce chest X-ray radiographs.…”
Section: Resultsmentioning
confidence: 50%
“…Table 6 shows a summary of this study’s LDRL values that they were lower than those of international organisations (EC 2018 ; IAEA 2004 ) and other previously published studies (Ahmed et al 2020 ; Baş Mor et al 2018 ; Hart et al 2012 ; Hoseini Motlagh et al 2020 ; Metaxas et al 2019 ; Mohsenzadeh et al 2018 ; Nyathi et al 2009 ; Suliman 2020 ). The reason for this is that the radiographers at the research site used the high kVp and low mAs technique to produce chest X-ray radiographs.…”
Section: Resultsmentioning
confidence: 50%
“…1dengan t p adalah ketebalan pasien, untuk ketebalan pasien pada pemeriksaan chest PA, abdomen PA, pelvic PA dan lumbar spine AP digunakan standar 23 cm dan 30 cm untuk semua proyeksi LAT. Sedangkah x adalah jarak antara detektor dengan holder detektor bagian muka yang diperkiran pada jarak 5 cm (Suliman, 2020).…”
Section: Materials Dan Metodeunclassified
“…Sehingga pada akhirnya dinamika nilai DRL bermuara pada kaidah As Low As Achievable Reasonable (ALARA). Nilai ESD hasil studi ini, jika dibandingkan dengan studi yang sejenis (Osibote and de Azevedo, 2008), (Suliman, 2020) serta nilai DRL Nasional (iDRL) (BAPETEN, 2019), serta acuan lembaga internasional seperti IAEA (IAEA., 2018) dan Europian Commission (EC) (European Union, 2014) ditampilkan pada Tabel 3.…”
Section: Pembahasanunclassified
“…3d dose delivery in radiology [13][14] and Radiotherapy 15 are essential to improve efficiency and avoid complications…”
Section: Iii-practical Use Of Dicom Imaging In Medicinementioning
confidence: 99%