It is necessary to measure the bounded-square output power (P1x1) for determination of soft tissue thermal index (TIS) for both scanning and non-scanning modes of an ultrasonic diagnostic equipment according to the new JIS (or IEC Standards). However, there is no report about how to measure P1x1 for an ultrasound probe with an inclined beam-axis (UPIB). We designed an angle-corrected 1 cm × 1 cm absorbing mask in accordance with the spirit of the IEC 62359 Ed.2.0 and established a reasonable process for measurement of TIS and MI of a paste type probe (PTP), which is a specially designed UPIB for examination of the blood flow in carotid artery more conveniently and efficiently. Both TIS and MI of the PTP were much less than 1.0 at all selectable presetting of the blood flow monitor, FURUHATA device, suggesting it’s as safe as a common ultrasonic Doppler diagnostic device.