“…Others 16,36/38 have reported that despite this deviation, lyophilized plasmas can be used to calculate the INR and to calibrate thromboplastins. Clearly, our study shows that calibration is more reliable with AK Cal plasmas than with ART Cal plasmas, 31 as also recently reported by Stevenson et al 39 However, AK Cal plasmas are limited in supply, costlier, and subject to ethical and regulatory issues and to possible batch-tobatch variability. Second, it is not easy to provide the wide range of INR values required for accurate ISI calibration on pooled AK Cal plasmas.…”