Evaluation of ion recombination and polarity effect on photon depth dose measurements using mini‐ and micro‐ion chamber
Antonella Fogliata,
Andrea Bresolin,
Pasqualina Gallo
et al.
Abstract:PurposeTo investigate the effect of ion recombination () and polarity () correction factors on percentage depth dose (PDD) curves for three ion chambers, using flat and flattening filter free (FFF) beams, across different broad field sizes. A method to assess these effects and their corresponding corrections is proposed.Methods and were evaluated following the IAEA TRS‐398 protocol for three ion chambers: PTW Semiflex‐3D‐31021, PinPoint‐3D‐31022, and Semiflex‐31010. PDD measurements were acquired from dmax to… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.