2021
DOI: 10.1051/radiopro/2021002
|View full text |Cite
|
Sign up to set email alerts
|

Evaluation of neutron spectra and dose equivalent from a Varian 2100C/D Medical Linear Accelerator: Monte Carlo simulation and a literature review

Abstract: In this study was carried out a review according to experimental and Monte Carlo studies in the literature on the neutron production from 18 MV, Varian 2100C/D linac. The effects of these neutrons were investigated on the total fluence, the energy spectra, and the dose equivalent. These factors were calculated as a function of depth and the radiation field size by simulation of linac head using of MCNPX2.6.0 code. The neutron strength was found equal to 1.23 × 1012 nGy−1.The results showed that with increasing… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1

Citation Types

0
3
0

Year Published

2021
2021
2024
2024

Publication Types

Select...
5

Relationship

0
5

Authors

Journals

citations
Cited by 5 publications
(3 citation statements)
references
References 37 publications
0
3
0
Order By: Relevance
“…It happens that the LNT model has been largely based on those Japanese bomb survivors, who therefore received an irradiation very different from what is received by radiotherapy patients. -The type of radiotherapy beams needs to be considered, knowing in particular the role played by particles with high relative biological effectiveness (RBE) such as neutrons in the Japanese survivors, while neutrons are not used any more in radiotherapy (Razghandi et al, 2021). -The shape of the risk curve at high doses is obviously the most important point to be considered when trying to evaluate the risk of the "very high" doses of radiotherapy.…”
Section: Reasons Of the Discrepancy Between Lnt Model Predictions And...mentioning
confidence: 99%
“…It happens that the LNT model has been largely based on those Japanese bomb survivors, who therefore received an irradiation very different from what is received by radiotherapy patients. -The type of radiotherapy beams needs to be considered, knowing in particular the role played by particles with high relative biological effectiveness (RBE) such as neutrons in the Japanese survivors, while neutrons are not used any more in radiotherapy (Razghandi et al, 2021). -The shape of the risk curve at high doses is obviously the most important point to be considered when trying to evaluate the risk of the "very high" doses of radiotherapy.…”
Section: Reasons Of the Discrepancy Between Lnt Model Predictions And...mentioning
confidence: 99%
“…Fast neutrons emitting from the accelerator head can directly reach the patient's bed or slow down by interaction with the components of the treatment room, including walls, tables, and the like, and eventually, be converted into thermal neutrons [18]. Therefore, a spectrum of neutron energy is expected anywhere on the patient's bed [19,20].…”
Section: Introductionmentioning
confidence: 99%
“…High-energy linac, i.e. 18 MV linac, generates high photoneutron contamination, with its amount varied between various field sizes [18,19]. Some studies confirm that photoneutrons generated inside the linac head are primarily in a fast neutron energy range.…”
Section: Introductionmentioning