BackgroundPercutaneous vertebroplasy (PVP) might lead to significant radiation exposure to patients, operators, and operating room personnel. Therefore, radiaton exposure is a concern. The aim of this study was to present a remote control cement delivery device and study whether it can reduce dose exposue to operators.MethodsAfter meticulous preoperative preparation, a series of 40 osteoporosis patients were treated with unilateral approach PVP using the new cement delivery divice. We compared levels of fluoroscopic exposure to operator standing on different places during operation. group A: operator stood about 4 meters away from X-ray tube behind the lead sheet. group B: operator stood adjacent to patient as using conventional manual cement delivery device.ResultsDuring whole operation process, radiation dose to the operator (group A) was 0.10 ± 0.03 (0.07-0.15) μSv, group B was 12.09 ± 4.67 (10–20) μSv. a difference that was found to be statistically significant (P < 0.001) between group A and group B.ConclusionNew cement delivery device plus meticulous preoperative preparation can significantly decrease radiation dose to operators.