Evaluation of signal-to-background and Mg II/Mg I ratios as response for the optimization of rare earth elements determination by inductively coupled plasma optical emission spectrometry

Abstract: Elementos terras raras (ETR) fazem parte de um grupo peculiar de elementos químicos e sua determinação em amostras geológicas é importante e complexa, devido às baixas concentrações e problemas com interferências espectrais e não espectrais. Neste trabalho, planejamentos experimentais foram utilizados para otimizar as condições operacionais de um espectrômetro de emissão óptica com plasma indutivamente acoplado (ICP OES) para a determinação de ETR. A razão entre o sinal analítico dos ETR e o sinal de fundo (SB… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
1
1
1

Citation Types

0
10
0

Year Published

2014
2014
2019
2019

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 13 publications
(10 citation statements)
references
References 36 publications
0
10
0
Order By: Relevance
“…In the present study, multivariate statistical methods were used for the optimization of the measuring parameters [39][40][41][42][43]. The robustness of the inductively coupled plasma was evaluated according to the Doehlert protocol, as described by Guimarães-Silva et al for ICP-OES [25]. In our case, the operating conditions were considered as factors, although the software of the MP-AES instrument allows to adjust only two parameters for the user: the read time and the nebulizing pressure.…”
Section: Discussionmentioning
confidence: 99%
See 3 more Smart Citations
“…In the present study, multivariate statistical methods were used for the optimization of the measuring parameters [39][40][41][42][43]. The robustness of the inductively coupled plasma was evaluated according to the Doehlert protocol, as described by Guimarães-Silva et al for ICP-OES [25]. In our case, the operating conditions were considered as factors, although the software of the MP-AES instrument allows to adjust only two parameters for the user: the read time and the nebulizing pressure.…”
Section: Discussionmentioning
confidence: 99%
“…To investigate the robustness of the microwave, plasma Mg was chosen as the test element on the same ionic and atomic lines that were described in earlier reports (MgII 280.271 nm/ MgI 285.213 nm) [24,25,37]. Although the MP-AES software offers less ionic lines compared to the ICPs, these two are both available.…”
Section: The Robustness Of the Microwave Plasma And The Threshold Conmentioning
confidence: 99%
See 2 more Smart Citations
“…The pressure of the nebulizer gas is inversely proportional to the sensitivity and Mg II /Mg I ratio that is in accordance with the literature. 34,[36][37][38] The speed of the peristaltic pump appears to be directly proportional to the sensitivity in Figure 2a but inversely proportional in Figure 2b. Figure 3 allows better visualization of the effects.…”
Section: Optimization Of Icp Oes Operationmentioning
confidence: 91%