2022
DOI: 10.1177/10406387221083621
|View full text |Cite
|
Sign up to set email alerts
|

Evaluation of the scil vCell 5, a novel laser- and impedance-based point-of-care hematology analyzer, for use in dogs and cats

Abstract: A novel laser- and impedance-based point-of-care hematology analyzer (POCA), the vCell 5 (scil Animal Care), providing a complete blood count with 5-part leukocyte differential count has recently been introduced to veterinary laboratories. We evaluated the analyzer for use in dogs and cats including method comparison and assessment of linearity, carryover, and precision. Fresh blood samples from 192 healthy and diseased dogs and 159 cats were analyzed, and results were compared to reference methods (i.e., micr… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3

Citation Types

0
8
0

Year Published

2023
2023
2023
2023

Publication Types

Select...
1

Relationship

1
0

Authors

Journals

citations
Cited by 1 publication
(8 citation statements)
references
References 37 publications
(122 reference statements)
0
8
0
Order By: Relevance
“…In addition to the impedance measuring principle for RBC and PLT counts, the scil POCA uses laser light flow cytometry for WBC differentiation providing a 5-part Diff. 62 Results of the Diff are presented both numerically in relative and absolute numbers, as well as graphically as a cytogram (dotplot) with cell size (low-angle scattered-light signals) shown on the y-axis and cell internal complexity (high-angle scattered-light signals) on the x-axis. The WBC populations visualized in the cytogram are NEU, LYM, MON, EOS, and BASO.…”
Section: Methodsmentioning
confidence: 99%
See 4 more Smart Citations
“…In addition to the impedance measuring principle for RBC and PLT counts, the scil POCA uses laser light flow cytometry for WBC differentiation providing a 5-part Diff. 62 Results of the Diff are presented both numerically in relative and absolute numbers, as well as graphically as a cytogram (dotplot) with cell size (low-angle scattered-light signals) shown on the y-axis and cell internal complexity (high-angle scattered-light signals) on the x-axis. The WBC populations visualized in the cytogram are NEU, LYM, MON, EOS, and BASO.…”
Section: Methodsmentioning
confidence: 99%
“… 31 We had determined intra-assay CV by 20 consecutive measurements performed from feline and canine K 3 -EDTA–anticoagulated blood samples with low, normal, and high WBC counts. 62 Briefly, the following CVs were obtained for a canine and feline sample with normal total WBC count and Diff (dog: NEU 2.43%, LYM 6.98%, MON 10.5%, EOS 11.2%; cat: NEU 1.9%, LYM 12.5%, MON 20.8%, EOS 31.0%).…”
Section: Methodsmentioning
confidence: 99%
See 3 more Smart Citations