The paper introduces the measuring principle of glass thermal expansion coefficient. It expresses the features and advantages of the linear variable differential transformer measuring principle by comparing the thermal expansion coefficient measurement methods analysis. Meanwhile, the paper introduces DIL402PC dilatometer measuring devices and measurement procedures. It use standard samples to execute thermal expansion instrument system check from the accuracy and repeatability. The result shows thermal expansion coefficient of linear expansion instrument bias of DIL402PC dilatometer is within the error range of the theoretical value. The process proves it has a wide range of applications.