Abstract:Overview of the sources of dc and ac ratio errors in conventional Cryogenic Current Comparator (CCC) is presented. Preliminary results of the measurements of ac ratio error in overlapped-tube type CCC are given. The measured maximum deviation from the nominal ratio in CCC used in MIRES is less than 2 ppm in frequency range 0-200 Hz.
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.