Failed TAVI in TAVI Implantation: TAVI Dislocation Followed by Ensuing Surgical Graft Resection

Abstract: We are presenting a case report of failed valve-in-valve treatment of severe aortic stenosis. A control ultrasonography after TAVI implantation revealed a severe aortic regurgitation of the graft which was subsequently unresolved with postimplantation dilatation. Second TAVI was implanted with cranial dislocation to the aortic root. Patient underwent a CT examination to clarify the TAVI in TAVI position. Patient underwent a surgical resection of TAVI with implantation of biological aortic valve prosthesis. In … Show more