2000
DOI: 10.1093/clinchem/46.7.1014
|View full text |Cite
|
Sign up to set email alerts
|

False-Positive Plasma Troponin I with the AxSYM Analyzer

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
3
2

Citation Types

1
6
0

Year Published

2002
2002
2018
2018

Publication Types

Select...
7
1

Relationship

0
8

Authors

Journals

citations
Cited by 15 publications
(7 citation statements)
references
References 1 publication
1
6
0
Order By: Relevance
“…These include heterophilic antibodies, 34 rheumatoid factor, 35 fibrin clots, 36 microparticles, 36 and malfunction of the analyzer itself. 37 The role of heterophilic antibodies in causing interference in immunoassays has been reported in a number of studies. 34,38 -41 The incidence of this interference varies considerably in the literature, 40 -43 ranging from 0.17 to 40%.…”
Section: False-positive Troponinmentioning
confidence: 99%
See 1 more Smart Citation
“…These include heterophilic antibodies, 34 rheumatoid factor, 35 fibrin clots, 36 microparticles, 36 and malfunction of the analyzer itself. 37 The role of heterophilic antibodies in causing interference in immunoassays has been reported in a number of studies. 34,38 -41 The incidence of this interference varies considerably in the literature, 40 -43 ranging from 0.17 to 40%.…”
Section: False-positive Troponinmentioning
confidence: 99%
“…Finally, malfunction of the analyzer itself is another reported cause of false elevations in cardiac TnI results. Galambos et al 37 reported several cases of falsely elevated plasma cardiac TnI levels due to a temporary malfunction of the AxSYM analyzer (Abbott Laboratories; Abbott Park, IL). The operation manual provided by the manufacturer mentions the possibility of a solution dispenser misalignment if the dispensers are repositioned incorrectly during the weekly maintenance.…”
Section: False-positive Troponinmentioning
confidence: 99%
“…Other endogenous substances such as rheumatoid factor, excess fibrin, and endogenous alkaline phosphatase can also produce false-positive results. [12][13][14] Our laboratory uses the AccuTnIC3 assay, which uses monoclonal antibodies specifically directed against human cardiac troponin I. The AccuTnIC assay uses mouse monoclonal antihuman cTnI alkaline phosphatase conjugate and uses a surfactant, bovine serum albumin matrix, and protein (bovine, goat, mouse) as reagents.…”
Section: Discussionmentioning
confidence: 99%
“…Despite the advantages of troponins, clinicians should keep in mind other processes that elevate troponin levels apart from myocardial infarction. 2 , 3 , 6 , 7 , 8 , 9 , 10 , 11 While some of the events with troponin elevation other than myocardial infarction are associated with myocardial injury, some of them occur as a result of troponin tests giving false positive results.…”
Section: Discussionmentioning
confidence: 99%
“…False positive troponin testing may result from: heterophile antibodies, Rheumatoid factor, or macroenzymes. Several additional examples of interfering substances are found in the literature; 3 , 10 , 11 among which are circulating antibodies from immunotherapies, vaccinations or blood transfusions, fibrin clots, immunocomplexes, and malfunction of the analyzers.…”
Section: Discussionmentioning