Abstract:The feasibility of simultaneous 1H/31P PET/MR imaging added-on to an existing 7T MRI system was assessed. Space constraints preclude using whole-body birdcages. Instead, an 8-channel 1H dipole array stacked with additional 8-31P channels is set up. By positioning PET detectors between the RF-shield and gradient coil, further shielding of the PET-crystals and electronics can be omitted. 8-channel 31P scans had a higher SNR compared to a birdcage setup. The 1H/31P setup gives ~5% degradation of PET sensitivity.… Show more
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.