2014
DOI: 10.1016/j.ancard.2014.08.009
|View full text |Cite
|
Sign up to set email alerts
|

FemoSeal Evaluation Registry (FER). Prospective study of femoral arterial closure with a mechanical system on 100 patients who underwent angioplasty procedures

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
10
0
1

Year Published

2018
2018
2024
2024

Publication Types

Select...
6

Relationship

0
6

Authors

Journals

citations
Cited by 9 publications
(12 citation statements)
references
References 14 publications
(20 reference statements)
1
10
0
1
Order By: Relevance
“…In the present series, the device was successfully deployed in 98.9% of cases, which is superior to the results from the CLOSE UP study (93.6%) 6 and ISAR CLOSURE trial (94.7%), 8 which could be explained by the 4 years of experience using this device at the present centre. These results are equivalent to those published by De Poli et al 9 in their registry which included only 100 patients. Deployment failure is usually associated with an increased risk of large haematomas, 11 but, no major adverse events occurred in the present series.…”
Section: Comparison With Femoseal Studiessupporting
confidence: 89%
See 1 more Smart Citation
“…In the present series, the device was successfully deployed in 98.9% of cases, which is superior to the results from the CLOSE UP study (93.6%) 6 and ISAR CLOSURE trial (94.7%), 8 which could be explained by the 4 years of experience using this device at the present centre. These results are equivalent to those published by De Poli et al 9 in their registry which included only 100 patients. Deployment failure is usually associated with an increased risk of large haematomas, 11 but, no major adverse events occurred in the present series.…”
Section: Comparison With Femoseal Studiessupporting
confidence: 89%
“…This study showed that VCDs were not inferior to manual compression in terms of vascular access site complications and reduced the time to haemostasis. The study of De Poli et al 9 was a single centre prospective registry conducted from November 2010 to April 2011, comparing the results of manual compression (n ¼ 111 patients), haemostatic bandage compression (n ¼ 43 patients), and FemoSeal mechanical closure (n ¼ 100 patients). In the study, the use of FemoSeal was effective in achieving haemostasis performed through a femoral arterial approach up to 7F and was associated with a very low rate of complications (one haematoma).…”
Section: Comparison With Femoseal Studiesmentioning
confidence: 99%
“…The safety and feasibility of this VCD on retrograde femoral artery access have been investigated and verified in previous studies [1][2][3][4][5][6][7][8]. A recent article showed that the device could also be safely used even for anterograde femoral artery access [9].…”
Section: Introductionmentioning
confidence: 74%
“…A r t i c l e compression was followed when technical failure was brought out in many reviews [1][2][3][4]6,7]. Even though the remarkable event has not been found after following manual compression in some studies [4][5][6][7], large hematomas were found in 36% of technical failure cases according to one large-scale survey, CLOSE-UP study [2]. A clear description of the result of technical failure cannot be found in the ISAR-CLOSURE study, another large-scale study [1].…”
Section: A C C E P T E Dmentioning
confidence: 99%
“…The safety and efficacy of this device, compared or not with MC, have been documented in several papers, 5,[9][10][11] but its use in antegrade common femoral access has not been yet investigated, although St Jude Medical does not advise against it in its "Instruction for use. "…”
Section: Introductionmentioning
confidence: 99%