Evaluate Acuros® XB dose calculation accuracy following TRS-483 recommendations in small static fields for flattened and un-flattened 6 MV X-ray beams.Methods: Field output factors were measured following TRS-483 recommendations using four radiation detectors. Two sets of field output factors were measured. One set was used to configure the beam model into Acuros® XB down to a jaw-defined field size of 1.0 cm × 1.0 cm. The second set was used to evaluate the differences between calculated and measured field output factors for MLC-fields down to a field size of 0.5 cm × 0.5 cm.Results: Acuros® XB showed an accuracy within 1.5% down to an MLC-field of 1.0 cm × 1.0 cm, for a focal spot size of 1.0 and 0.0 mm in the cross and in-plane directions. For an MLC-field of 0.5 cm × 0.5 cm, an agreement was found within 3% between calculated and measured field output factors. These results were addressed by optimizing the focal spot size to minimize the differences between calculated and measured dose profiles.Conclusions: By optimizing the focal spot size, Acuros® XB showed an acceptable agreement within 3% down to an MLC-field of 0.5 cm × 0.5 cm. The results of this work suggest that if static and modulated delivery of very small targets is planned, then a field output factor table down to a field size of 1.0 cm is required in the beam configuration model.