2020
DOI: 10.1136/neurintsurg-2020-016115
|View full text |Cite
|
Sign up to set email alerts
|

First clinical multicenter experience with the new Scepter Mini microballoon catheter

Abstract: BackgroundBalloon-assisted techniques can improve the endovascular treatment of cerebrospinal vascular malformations. The aim of this study was to report the first clinical multicenter experience with the new Scepter Mini dual-lumen microballoon catheter.MethodsPatients with cerebral or spinal vascular malformations treated with the Scepter Mini at seven European neurovascular centers were retrospectively reviewed. Clinical data, angiographic features of the vascular malformations, procedural parameters includ… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
4
1

Citation Types

1
39
0

Year Published

2020
2020
2023
2023

Publication Types

Select...
7
1

Relationship

1
7

Authors

Journals

citations
Cited by 37 publications
(40 citation statements)
references
References 16 publications
1
39
0
Order By: Relevance
“…A second study, describing the use of the Scepter Mini balloon microcatheter, described 34 instances of balloon use in 20 patients across multiple centers. 9 The mean diameter of the catheterized blood vessel was 1.9 mm with technical failures occurring in 4/23 embolizations. 9 There were no complications directly related to the Scepter Mini, and the technical failures were all due to insufficient stability of the balloon and distal catheter tip within the target vessel.…”
Section: Discussionmentioning
confidence: 95%
See 2 more Smart Citations
“…A second study, describing the use of the Scepter Mini balloon microcatheter, described 34 instances of balloon use in 20 patients across multiple centers. 9 The mean diameter of the catheterized blood vessel was 1.9 mm with technical failures occurring in 4/23 embolizations. 9 There were no complications directly related to the Scepter Mini, and the technical failures were all due to insufficient stability of the balloon and distal catheter tip within the target vessel.…”
Section: Discussionmentioning
confidence: 95%
“…9 There were no complications directly related to the Scepter Mini, and the technical failures were all due to insufficient stability of the balloon and distal catheter tip within the target vessel. 9 The final study looked at the use of balloons in a spinal vascular malformation. 11 In our series, there were no technical failures of the balloon.…”
Section: Discussionmentioning
confidence: 99%
See 1 more Smart Citation
“…The use of this catheter has only been described three other times in the literature. [9][10][11] This is the first described use in a VOGM.…”
Section: Discussionmentioning
confidence: 99%
“…5 , 15 , 16 A recent study also describes use of SMB in brain arteriovenous malformations. 17 Utilizing dual lumen balloon microcatheters for liquid embolic injection has been shown to be associated with reduced fluoroscopic times and better penetration of targeted nidus with liquid embolic system, as well as lower instances of microcatheter entrapment. 7 , 18 However, the majority of vascular pedicles that are embolized in cerebrovascular malformations are less than 3 mm in size.…”
Section: Discussionmentioning
confidence: 99%