1996
DOI: 10.1017/s0022215100135777
|View full text |Cite
|
Sign up to set email alerts
|

Fractured tracheostomy tube: an overlooked foreign body

Abstract: A rare case of an overlooked fractured metallic Fuller's tracheostomy tube in the posterior basal segment of the left lung is reported along with a review of the literature. Incidence, aetiology, delay in diagnosis, complications and management of overlooked foreign bodies are discussed. Various factors contributing to the fracture of tracheostomy tubes are reviewed.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
5

Citation Types

0
19
0

Year Published

2000
2000
2020
2020

Publication Types

Select...
10

Relationship

0
10

Authors

Journals

citations
Cited by 28 publications
(19 citation statements)
references
References 11 publications
0
19
0
Order By: Relevance
“…Prolonged wear, ageing of the tubes and repeated sterilization have been proposed as risk factors of a fractured tracheostomy tube [8,11-14]. Alkaline bronchial secretion, tissue reactivity from plastic tubes, long continued high internal stresses on the surface and manufacturing defects were also reported as causes of this complication [11-13]. In our opinion, the fracture of the tracheostomy tube in this patient may have been due to prolonged wear and ageing of the tube.…”
Section: Discussionmentioning
confidence: 99%
“…Prolonged wear, ageing of the tubes and repeated sterilization have been proposed as risk factors of a fractured tracheostomy tube [8,11-14]. Alkaline bronchial secretion, tissue reactivity from plastic tubes, long continued high internal stresses on the surface and manufacturing defects were also reported as causes of this complication [11-13]. In our opinion, the fracture of the tracheostomy tube in this patient may have been due to prolonged wear and ageing of the tube.…”
Section: Discussionmentioning
confidence: 99%
“…1 5 6 In this case, it was at the junction of the tube and neck plate. Some of the mechanisms proposed include prolonged usage leading to increased wear and tear, accumulation of alkaline tracheobronchial secretion leading to softening and weakening of the tube, and manufacturing defects 2–4 7 8. In the above mentioned case, the child had undergone tracheostomy very early in life, at around 7 months of age, and it had been retained until 6 years of age.…”
Section: Discussionmentioning
confidence: 99%
“…7 A series of nine cases over a period of 8 years has been reported by Gupta in 1996. 8 Incidence of broken tracheostomy tube that remained undetected for a long period has been rare. Sreenath and Mahendrakar described a case of broken tracheostomy tube that was removed tube four years after inhalation.…”
Section: Discussionmentioning