“…On the contrary, semi-volatile CWAs such as VX has a low vapor pressure, potentially causing an insufficient collection of its vapor sample. However, it is important to be able to collect semi-volatile CWA samples, from the ambient atmosphere, in order to analyze them using the highly sensitive GC/MS instruments at off-site laboratories; previous endeavors to address this issue have been reported [14,19,21]. Herein, we investigate the elution behaviors of five CWAs (GB, GD, tabun (GA), VX, and HD) on several resins by monitoring their mass spectrometric response to identify adsorbent resins that are effective in trapping volatile and semi-volatile CWAs (nerve agents and HD) in air samples.…”