Gebauer SLc Original and Moria One‐Use Plus automated microkeratomes for ultrathin Descemet's stripping automated endothelial keratoplasty preparation

Abstract: The SLc and the OUP system are both suited for the preparation of UT-DSAEK lamellae. Neither system differed significantly in variability, LSR or ECL, which did not increase during a 6-day incubation period.

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1
1

Citation Types

0
6
0

Year Published

2017
2017
2021
2021

Publication Types

Select...
7

Relationship

0
7

Authors

Journals

citations
Cited by 9 publications
(6 citation statements)
references
References 34 publications
0
6
0
Order By: Relevance
“…These results were similar to those reported by Fuest et al . [ 11 12 ] except that the endothelial keratoplasty lenticules in their series were dissected only from corneoscleral buttons, not from fresh whole eyes. The superiority of the SLc system over OUP in this regard may be due to the use of a transparent visual applanation plate in SLc for setting the desired cutting diameter.…”
Section: Discussionmentioning
confidence: 99%
See 3 more Smart Citations
“…These results were similar to those reported by Fuest et al . [ 11 12 ] except that the endothelial keratoplasty lenticules in their series were dissected only from corneoscleral buttons, not from fresh whole eyes. The superiority of the SLc system over OUP in this regard may be due to the use of a transparent visual applanation plate in SLc for setting the desired cutting diameter.…”
Section: Discussionmentioning
confidence: 99%
“…[ 9 10 ] Hand-guided Moria Carriazo-Barraquer (CBm) and fully automatic systems such as Gebauer SLc Original (SLc) and Moria One-Use Plus (OUP) are suitable systems that are being used for the standard preparation of DSAEK tissues. [ 11 12 ] Although multiple factors can influence the thickness of provided DSAEK lamellae,[ 13 14 15 ] the SLc system excepting its more agreement with the selected cutting depth was comparable with the CBm and OUP in terms of variability, lamellar surface roughness, or endothelial cell loss. [ 11 12 ]…”
mentioning
confidence: 99%
See 2 more Smart Citations
“…The most common method is a subjective roughness score applied to scanning electronic microscopy (SEM) [10,[14][15][16][17]. The next most common is the objective measure provided by atomic force microscopy (AFM) [8,9,11,18].…”
Section: Introductionmentioning
confidence: 99%