We proposed a simple visual method for evaluating the dynamic tumor tracking (DTT) accuracy of a gimbal mechanism using a light field. A single photon beam was set with a field size of 30×30 mm2 at a gantry angle of 90°. The center of a cube phantom was set up at the isocenter of a motion table, and 4D modeling was performed based on the tumor and infrared (IR) marker motion. After 4D modeling, the cube phantom was replaced with a sheet of paper, which was placed perpendicularly, and a light field was projected on the sheet of paper. The light field was recorded using a web camera in a treatment room that was as dark as possible. Calculated images from each image obtained using the camera were summed to compose a total summation image. Sinusoidal motion sequences were produced by moving the phantom with a fixed amplitude of 20 mm and different breathing periods of 2, 4, 6, and 8 s. The light field was projected on the sheet of paper under three conditions: with the moving phantom and DTT based on the motion of the phantom, with the moving phantom and non‐DTT, and with a stationary phantom for comparison. The values of tracking errors using the light field were 1.12±0.72, 0.31±0.19, 0.27±0.12, and 0.15±0.09 mm for breathing periods of 2, 4, 6, and 8 s, respectively. The tracking accuracy showed dependence on the breathing period. We proposed a simple quality assurance (QA) process for the tracking accuracy of a gimbal mechanism system using a light field and web camera. Our method can assess the tracking accuracy using a light field without irradiation and clearly visualize distributions like film dosimetry.PACS number(s): 87.56 Fc, 87.55.Qr