Gonad shielding in pelvic radiography: modern optimised X-ray systems might allow its discontinuation

Abstract: Objective: As gonad shielding is currently under debate, this study evaluates the practice, from its introduction in about 1905 until today. Methods: The literature was searched for developments in shielding and insights into the effects of ionising radiation on gonads. Based on own pre-1927 dose reconstructions, reported doses after 1927, a 2015-report from the European Union and recent own measurements, the effects of technological evolution and optimisation on radiation dose and hereditary risk were assesse… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

3
17
1
2

Year Published

2020
2020
2024
2024

Publication Types

Select...
5
1

Relationship

0
6

Authors

Journals

citations
Cited by 26 publications
(24 citation statements)
references
References 48 publications
3
17
1
2
Order By: Relevance
“…The K a,i values of this study for the standard and added beam filtrations, 1.6 and 1.2 mGy respectively, are more than double the K a,i values of 0.5-0.6 mGy reported by Jeukens, et al 8 in 2018. The more than double K a,i values in our study are due in part to three differences: (a) The SID of the X-ray Source description A previously developed code for simulating CT source was used by making the tube stationary and adjusting other beam parameters for digital radiography defined in Table II Energy spectra: 1 keV bins from 1-150 keV generated using Spektr3.0 for each tube potential.…”
Section: Discussioncontrasting
confidence: 42%
See 2 more Smart Citations
“…The K a,i values of this study for the standard and added beam filtrations, 1.6 and 1.2 mGy respectively, are more than double the K a,i values of 0.5-0.6 mGy reported by Jeukens, et al 8 in 2018. The more than double K a,i values in our study are due in part to three differences: (a) The SID of the X-ray Source description A previously developed code for simulating CT source was used by making the tube stationary and adjusting other beam parameters for digital radiography defined in Table II Energy spectra: 1 keV bins from 1-150 keV generated using Spektr3.0 for each tube potential.…”
Section: Discussioncontrasting
confidence: 42%
“…2,7 Due to technical improvements, current clinical radiation doses without gonadal shielding have been estimated to be less than 5% of the values 7 decades ago. 8 In 2018, K a,e values were reported in a European survey of abdominal X-rays of adults of 0.8 AE 0.5 mGy. 8 Despite its stature as a 'gold standard" for more than 50 yr, 9 the practice of gonadal shielding began to be questioned in the last decade as additional data was published.…”
Section: Introductionmentioning
confidence: 99%
See 1 more Smart Citation
“…Para los hombres, puede haber un beneficio marginal, pero los posibles efectos secundarios negativos hacen cuestionar su uso. La interrupción de la protección de las gónadas, por lo tanto, parece justificable en el momento actual (20) . Esto es algo que la profesión radiológica debería tomar en serio dada la orientación y la legislación basados en datos de larga data.…”
Section: Discussionunclassified
“…La evolución tecnológica y la optimización han sido capaces de reducir las dosis de gónadas al 0.5-2% de los valores que se generaban a principios del siglo pasado (aprox. 1905) (20) .…”
Section: Dosis De Radiaciónunclassified