High‐field downfield MR spectroscopic imaging in the human brain
İpek Özdemir,
Semra Etyemez,
Peter B. Barker
Abstract:PurposeTo investigate the feasibility of downfield MR spectroscopic imaging (DF‐MRSI) in the human brain at 7T.MethodsA 7T DF‐MRSI pulse sequence was implemented based on the previously described methodology at 3T, with 3D phase‐encoding, spectral‐spatial excitation, and frequency selective refocusing. Data were pre‐processed followed by analysis using the “LCModel” software package, and metabolite maps created from the LCModel results. Total scan time, including brain MRI and a water‐reference MRSI, was 24 m… Show more
Set email alert for when this publication receives citations?
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.