High sensitivity HPLC method for analysis of in vivo extracellular GABA using optimized fluorescence parameters for o -phthalaldehyde (OPA)/sulfite derivatives

Abstract: Reversed-phase HPLC with derivatization using o-phthalaldehyde (OPA) and sulfite allows electrochemical detection of γ-aminobutyric acid (GABA) in microdialysis samples. However, OPA/sulfite derivatives have been reported to produce lower fluorescent yield than OPA derivatives using organic thiols as the nucleophile. To overcome this limitation we examined excitation and emission spectra, reaction time, pH, and concentration of reagents in the derivatization solution. Optimal detection parameters were determin… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
2
2
1

Citation Types

1
16
0

Year Published

2018
2018
2024
2024

Publication Types

Select...
10

Relationship

0
10

Authors

Journals

citations
Cited by 23 publications
(17 citation statements)
references
References 37 publications
1
16
0
Order By: Relevance
“…Using 3D Monte Carlo reactiondiffusion simulations, Scimemi (Scimemi, 2014) estimated a steady-state GABA concentration of 0.1-1 µM in the cleft. This is similar to the concentration of measured in the cerebrospinal fluid (de Groote and Linthorst, 2007;Lerma et al, 1986;Zandy et al, 2017). Using the activation and desensitization parameters presented above, we estimate that in the presence of 0.1 µM GABA, 0.05% of receptors are in the active state and 0.2% in the desensitized state.…”
Section: Discussionsupporting
confidence: 82%
“…Using 3D Monte Carlo reactiondiffusion simulations, Scimemi (Scimemi, 2014) estimated a steady-state GABA concentration of 0.1-1 µM in the cleft. This is similar to the concentration of measured in the cerebrospinal fluid (de Groote and Linthorst, 2007;Lerma et al, 1986;Zandy et al, 2017). Using the activation and desensitization parameters presented above, we estimate that in the presence of 0.1 µM GABA, 0.05% of receptors are in the active state and 0.2% in the desensitized state.…”
Section: Discussionsupporting
confidence: 82%
“…Fractions collected from islet perifusion in the BioRep islet perifusion device or supernatants from static incubations in KRBH buffer were analyzed for GABA content using an EICOM HTEC-500 HPLC-ECD with autosampler, online automated OPA-derivatization, and Eicompak FA-3ODS separation column. This automated detection technique is linearly sensitive for GABA from the nano-molar to milli-molar range 107 . Insulin content in the same sample fraction was determined by ELISA kit (Mercodia).…”
Section: Detection Of Gaba By Hplc With Electrochemical Detectormentioning
confidence: 99%
“…Thus, the use of fluorescence detection makes it possible to avoid the co-elution that occurs so often with UV detection. There are many examples of the use of fluorescence detection in HPLC analysis, and while there are some that include florescent test chemicals (Cecil and Rutan, 1991), most involve the fluorescence derivatization of amino acids, amino acid derivatives, or other non-fluorescent materials (Purcell et al, 1976;Mashat et al, 2008;Zandy et al, 2017;Muramoto and Kamiyama, 1985;Tapuhi et al, 1981;Imai and Watanabe, 1981). Like ADRA and DPRA, the High Throughput with Dansyl Cysteamine (HTS-DCYA) method is an in chemico test that was developed as an alternative to animal testing for skin sensitization.…”
Section: Introductionmentioning
confidence: 99%