1992
DOI: 10.1007/bf02368172
|View full text |Cite
|
Sign up to set email alerts
|

Hoffmann half-frame external fixation rigidity and its relationship to universal joint slippage

Abstract: Hoffmann half-frame external fixation device configurations often fail under minimal loads secondary to joint slippage. In these experiments improved universal joints that were developed in an earlier study were tested on Hoffmann half-frame assemblies. The rigidity of selected half-frame configurations was tested in four modes (axial compression, torsion, medial-lateral and anterior-posterior four-point bending). These results were compared to those of an earlier, similar study looking at the standard Hoffman… Show more

Help me understand this report

Search citation statements

Order By: Relevance

Paper Sections

Select...
1

Citation Types

0
1
0

Year Published

1997
1997
2016
2016

Publication Types

Select...
4

Relationship

0
4

Authors

Journals

citations
Cited by 4 publications
(1 citation statement)
references
References 21 publications
0
1
0
Order By: Relevance
“…Basically, frame design or construct requires a combination of wires, pins, rods, clamps and rings to ultimately assemble a stable construct [3,14,30]. The various pins including transfixion wire, tapered cortical half-pin, trocar-tip tapered half-pin, self-drilling tapered, self-drilling cylindrical and cancellous tapered [14,[33][34][35][36]. To be noticed, pin insertion technique is important to be chosen by surgeons as it can improve the initial pin torque resistance in order to minimize pin loosening [37].…”
Section: General Features Of the External Fixator Configurationsmentioning
confidence: 99%
“…Basically, frame design or construct requires a combination of wires, pins, rods, clamps and rings to ultimately assemble a stable construct [3,14,30]. The various pins including transfixion wire, tapered cortical half-pin, trocar-tip tapered half-pin, self-drilling tapered, self-drilling cylindrical and cancellous tapered [14,[33][34][35][36]. To be noticed, pin insertion technique is important to be chosen by surgeons as it can improve the initial pin torque resistance in order to minimize pin loosening [37].…”
Section: General Features Of the External Fixator Configurationsmentioning
confidence: 99%