ObjectiveNickel allergy is common; endovascular specialists are often confronted with nickel allergic patients ahead of the implantation of endovascular devices, many of which are nickel-containing. Our aim was to elucidate whether nickel hypersensitivity is significantly associated with worse or adverse outcomes after placement of a nickel-containing endovascular device.MethodsInclusion criteria were: endovascular and transcatheter procedures for coronary, structural heart, neurovascular and peripheral vascular pathology involving nickel-allergic patients. All adverse outcomes were included as defined by included studies. A systematic review and meta-analysis were undertaken using a random-effects model. Searches of MEDLINE and EMBASE were conducted for articles published 1947–2019.Results190 records were identified, 78 articles were included for qualitative synthesis and 15 met criteria for meta-analysis. Patch-test confirmed nickel allergy was associated with an increased risk of adverse outcomes following implantation of a nickel-containing endovascular device (n=14 articles, 1740 patients; OR 2.61, 95% CI 1.41 to 4.85). This finding further was observed in coronary (n=12 articles, 1624 patients; OR 1.94, 95% CI 1.16 to 3.23) and structural heart subgroups (n=2 articles, 83 patients; OR 52.28, 95% CI 1.31 to 2079.14), but not in the neurovascular subgroup (n=1 article, 33 patients; OR 3.04, 95% CI 0.59 to 15.72) or with a patient-reported history of nickel allergy (n=2 articles, 207 patients; OR 2.14, 95% CI 0.23 to 19.70).ConclusionsPatch-tested nickel allergy is associated with an increased risk of adverse outcomes following endovascular device implantation and alternative treatment options should be considered. Specialists faced with patients’ self-reporting nickel allergy should consider proceeding to diagnostic patch-testing.