The determination of solid fat content (SFC) is an important analytical procedure in the food industry. The most common way to determine SFC is by low-resolution pulsed NMR (p-NMR). Although this technique is very sensitive and easy to use, it has the disadvantage that it cannot be used for on-line measurements. The present work compares new technologies to determine SFC on-line. On-line ultrasonic spectroscopy and NMR-MOUSE (NMR mobile universal surface explorer) techniques were compared with off-line p-NMR measurements and there was a good correlation between the values obtained. Ultrasonic measurements accurately described the SFC variation, whereas NMR-MOUSE determinations need to be improved to some extent owing to a strong temperature and motion dependence. These two techniques can be used as on-line methodologies to determine SFC during the crystallization of fats. FIG. 3. (a) Calibration curve for the NMR-MOUSE technology showing the linear relationship between the NMR-MOUSE reading and the conventional pulsed NMR (p-NMR) technology, mq20 (shown as liquid percentage). (b) Comparison of SFC values of different dilutions of sample A measured by traditional p-NMR, mq20, and NMR-MOUSE (x and L L, respectively). For abbreviations see Figure 1. SOLID FAT CONTENT DETERMINATION ON-LINE 317 JAOCS, Vol. 82, no. 5 (2005) FIG. 5. Comparison between SFC values measured with traditional p-NMR, mq20 (line with no symbols), ultrasonics (L L), and NMR-MOUSE (L) for sample A (a) and B (b), diluted with 50% of canola oil crystallized at 30°C. For abbreviations see Figures 1 and 3.